COVID-19 & Flu A/B Ag Result Interpreation

2020-12-11 11:31Views:157times

[仅供海外市场]


Positive Result:

Covid-19 antigen positive result: Both the Test "T" position and Control "C" position appear visible color lines.

Influenza A/B positive result: Test "A" position, test B” position and Control "C" position appear visible color lines.

Influenza A positive result: Test "A" position and Control "C" position appear visible color lines.

Influenza B positive result: Test "B" position and Control "C" position appear visible color lines. Note: If the test position shows a faint color, the result is also positive.


Negative Result:Only the two Control "C" positions appear color, while other test positions are all blank.


Invalid Result: If the control "C" remains blank without color appearing, the test result is invalid because the detecting function is failed to work. Thus, retest the specimen is required.  


For Covid-19 Antigen results:

                                                                   

Positive                                     Negative                                                 Invalid


For Influenza A/B antigen results:

         


Influenza A&B   Influenza A     Influenza B        Negative                                                     Invalid

    positive           positive            positive

                               